液相色谱-串联质谱法测定原发性肝癌患者血清中7种甾类激素
CSTR:
作者:
作者单位:

第二军医大学东方肝胆外科医院药材科,第二军医大学东方肝胆外科医院药材科,第二军医大学东方肝胆外科医院药材科,第二军医大学东方肝胆外科医院药材科,第二军医大学东方肝胆外科医院药材科,第二军医大学东方肝胆外科医院药材科,第二军医大学东方肝胆外科医院药材科

作者简介:

通讯作者:

中图分类号:

基金项目:


LC-MS/MS for determination of seven steroid hormones in sera of primary hepatocellular carcinoma patients
Author:
Affiliation:

Department of Pharmacy,Eastern Hepatobiliary Surgery Hospital,Second Military Medical University,Department of Pharmacy,Eastern Hepatobiliary Surgery Hospital,Second Military Medical University,Department of Pharmacy,Eastern Hepatobiliary Surgery Hospital,Second Military Medical University,Department of Pharmacy,Eastern Hepatobiliary Surgery Hospital,Second Military Medical University,Department of Pharmacy,Eastern Hepatobiliary Surgery Hospital,Second Military Medical University,Department of Pharmacy,Eastern Hepatobiliary Surgery Hospital,Second Military Medical University,Department of Pharmacy,Eastern Hepatobiliary Surgery Hospital,Second Military Medical University

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的 建立高效液相色谱-串联质谱(LC-MS/MS)法,用于测定原发性肝癌患者血清中7种甾类激素的含量。方法 血清样品采用乙酸乙酯液-液萃取,经丹磺酰氯衍生化后测定。色谱柱为Agilent Poroshell 120 EC-C18柱(2.1 mm×150 mm,2.7μm),流动相为乙腈-0.1%(VV)甲酸水溶液,梯度洗脱,流速为0.3 mL/min,柱温为35℃,质谱检测采用动态多反应离子监测模式(MRM)。结果 7种甾类激素(雌二醇、雌三醇、雌酮、氢化可的松、睾酮、雄烯二酮和黄体酮)在8 min内达到基线分离。在相应浓度范围内线性关系良好,相关系数均>0.99,日内及日间精密度RSD均<15%,提取回收率为80%~119%,基质效应为85%~112%,样品在放置6 h后处理、处理后室温放置24 h、3个冻融循环以及-80℃保存30 d等测定过程中均稳定性良好。结论 该方法操作简便、结果准确、灵敏度高、重现性好,适用于原发性肝癌患者体内甾类激素水平检测。

    Abstract:

    Objective To establish a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for determination of seven steroid hormones in sera of patients with primary hepatocellular carcinoma. Methods The serum samples were extracted with ethyl acetate, and then derivatized with dansyl chloride. Chromatographic column was Agilent Poroshell 120 EC-C18 (2.1 mm×150 mm, 2.7 μm) column, mobile phase was acetonitrile and 0.1% formic acid-water solution, flow rate was 0.3 mL/min, the column temperature was 35℃, and the MS detection was selected in dynamic MRM mode. Results Seven steroid hormones (estradiol, estriol, estrone, hydrocortisone, testosterone, androstenedione and progesterone) were baseline separated within 8 min, which possessed good linear relationship (r>0.99) within the linear concentration range; the intra-day and inter-day precisions were less than 15%, the recoveries were ranged from 80% to 119%, the matrix effect was 85%-112%. The detection results were stable when the samples were examined 6 h after collection, 24 h after treatment at room temperature, after three freeze-thaw cycles, and 30 days when stored at -80℃. Conclusion The current LC-MS/MS method is simple, accurate, sensitive and reproducible, and it can be applied to determine steroid hormones in the sera of patients with primary hepatocellular carcinoma.

    参考文献
    相似文献
    引证文献
相关视频

分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2016-02-19
  • 最后修改日期:2016-03-19
  • 录用日期:2016-06-27
  • 在线发布日期: 2016-07-27
  • 出版日期:
文章二维码
重要通知
友情提醒: 近日发现论文正式见刊或网络首发后,有人冒充我刊编辑部名义给作者发邮件,要求添加微信,此系诈骗行为!可致电编辑部核实:021-81870792。
            《海军军医大学学报》编辑部
关闭