【打印本页】 【下载PDF全文】 【HTML】 查看/发表评论下载PDF阅读器关闭

←前一篇|后一篇→

过刊浏览    高级检索

本文已被:浏览 1761次   下载 2067 本文二维码信息
码上扫一扫!
新型左心耳封堵器封堵犬左心耳的实验研究
储国俊1,2,张志钢3,巫俊2,阚通1,张莎1,赵仙先1,吴弘1*
0
(1. 第二军医大学长海医院心血管内科, 上海 200433;
2. 中国人民解放军镇江船艇学院门诊部, 镇江 212003;
3. 南京军区福州总医院心内科, 福州 350001
*通信作者)
摘要:
目的 采用国产新型镍钛合金双盘状镍钛合金封堵器封堵犬左心耳,评估该封堵器的可行性、安全性及生物相容性。方法 选取12只健康杂种犬,经股静脉途径穿刺房间隔封堵实验犬左心耳,于术后即刻和术后1、3、6个月通过心电图监测、房间隔穿刺造影、经食管超声、病理检查等方法检测封堵器的位置和植入效果。结果 术中即刻12只实验犬中11例成功植入封堵器,1例术后发生封堵器脱落。2例因术中封堵器大小不合适,回收后更换合适封堵器成功植入。术后即刻经食管超声和左房造影检查可见封堵器形态、位置良好,3例实验犬左心耳封堵后出现少量分流,余实验犬左心耳被完全封堵。术后1、3、6个月时造影结果可见封堵器无移位,无残余分流,且对冠脉血流无影响。术后6个月经食管超声复查和病理检查可见左心耳被完全封堵,封堵器表面与左心房内均无血栓形成,肝、肾、脾无栓塞或梗死灶,且封堵器表面完全被内皮细胞覆盖。结论 国产新型镍钛合金封堵器具有良好的生物相容性,使用该封堵器行经导管左心耳封堵术切实可行,中期效果理想。
关键词:  左心耳  封堵器    心房颤动
DOI:10.16781/j.0258-879x.2016.08.0943
投稿时间:2016-05-25修订日期:2016-07-04
基金项目:上海市自然科学基金(13ZR1409500).
A novel occluder for occluding the left atrial appendage: an experimental study with canines
CHU Guo-jun1,2,ZHANG Zhi-gang3,WU Jun2,KAN Tong1,ZHANG Sha1,ZHAO Xian-xian1,WU Hong1*
(1. Department of Cardiology, Changhai Hospital, Second Military Medical University, Shanghai 200433, China;
2. Department of Outpatient, Zhenjiang Watercraft College of PLA, Zhenjiang 212003, Jiangsu, China;
3. Department of Cardiology, Fuzhou General Hospital, PLA Nanjing Military Area Command, Fuzhou 350001, Fujian, China
*Corresponding author)
Abstract:
Objective To adopt a novel domestic double-disc nitinol occluder to occlude the left atrial appendage of canines and evaluate its feasibility,safety and biocompatibility in experimental canine model. Methods Totally 12 healthy dogs were used in this study; their atrial septal was punctured to occlude the left atial appendage via the femoral vein. The position and effect of the implanted occluders were evaluated by electrocardiogram (ECG) monitoring, atrial septal puncture angiography, transesophageal echocardiography (TEE), and pathological examination immediately, 1, 3, and 6 months after the operation. Results Among the 12 dogs, 11 were successfully implanted with the occluders immediately after operation, with the occluder in 1 case dropped after the operation. Two cases were firstly implanted with occluders of inappropriate size, which were withdrawn and successfully replaced by appropriate ones. The shape and location of the occluders were good as shown by TEE and left atrial angiography examination, with 3 cases having a little shunt and other cases completely occluded immediately after the operation as shown by TEE and left atrial angiography examination. There was no occluder shift, residual shunt or influence on coronary blood flow at 1, 3, and 6 months after the operation. It was shown that the left atrial appendage was completely blocked at 6 months after operation, and there was no thrombus formation on the surface of occluders or in the left atrium; there was no embolization or infarction in the liver,kidney or spleen, with the surface of the occluders completely covered with endothelial cells. Conclusion The novel self-made nitinol occluder has good biological compatibility, making it feasible to occlude the left atrial appendage via transcatheter approach, with satisfactory middle-term effects.
Key words:  left atrial appendage  occluder  canine  atrial fibrillation