Application of three automatic blood group testing instruments in cross matching test
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

Supported by Key Public Health Discipline Construction Program of Shanghai (12GWZX0202).

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective To apply three automatic blood group testing instruments, WADianaR○ (Diana), Techno TwinStation and ORTHO AutoVue Innova/Ultra, in clinical cross match test and to assess their effect, so as to ensure safe clinical blood transfusion. Methods The three instruments were used for cross matching test of 8 073 blood recipients and blood specimens of Shanghai Blood Center. The incompatible positive samples were reexamined by experiment cross matching antiglobulin test tube method. Results The 3 instruments detected 589 positive blood samples. There was no false positive result in the WADianaR○(Diana) system; Techno TwinStation and ORTHO AutoVue Innova/Ultra analyzer had 6 and 19 false positive results, respectively. Conclusion All the three instruments can accurately detect the positive samples in clinical cross match test, and WADianaR○ system has a more prominent performance.

    Reference
    Related
    Cited by
Related Videos

Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:July 31,2012
  • Revised:November 07,2012
  • Adopted:
  • Online: January 22,2013
  • Published: