Performance of PROPELLER technique in reducing susceptibility artifacts in diffusion-weighted MR images
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    ObjectiveTo evaluate the performance of PROPELLER diffusion-weighted magnetic resonance image (DWI) in reducing image distortion caused by susceptibility artifacts. MethodsForty volunteers with metal denture were enrolled in the present study; their PROPELLER and echo-planar DWI were taken for the skull base. The quality of each image was evaluated by using a four-point scoring system. Qualitative comparison of image distortion between PROPELLER DWI and echo-planar DWI was performed by ROC analysis. ResultsThe area under the ROC curve was 0.979 (P<0.01), indicating a significant difference in the image distortion between PROPELLER DWI and echo-planar DWI. PROPELLER technique could greatly reduce the image distortion associated with magnetic susceptibility artifacts. ConclusionPROPELLER technique can greatly reduce metal denture-induced susceptibility-related image distortion, and ROC curve can be used for quantitative analysis of subjective assessment of image quality.