【打印本页】 【下载PDF全文】 【HTML】 查看/发表评论下载PDF阅读器关闭

←前一篇|后一篇→

过刊浏览    高级检索

本文已被:浏览 2419次   下载 2015 本文二维码信息
码上扫一扫!
PROPELLER技术对磁共振弥散加权成像中磁敏感伪影改善的评价
刘小艳1,刘婷婷1,王骏2*
0
(1. 南通大学附属医院影像科,南通 226001
2. 南京军区南京总医院影像科,南京 210002
*通信作者)
摘要:
目的评价PROPELLER技术对磁共振弥散加权成像(DWI)中磁敏感伪影的改善。方法对40例含有金属义齿的志愿者颅底区域采用PROPELLER技术和平面回波成像(EPI)技术进行DWI成像。采用四点评分系统对图像质量进行评估,并对结果行ROC曲线分析。结果ROC曲线下面积(AUC)=0.979(P<0.01),表明金属义齿者颅底的PROPELLER DWI 成像技术与EPI DWI成像技术所产生的磁敏感伪影差异有统计学意义,前者能明显减少磁敏感伪影。结论 PROPELLER DWI成像技术能显著改善金属义齿所致的磁敏感伪影,利用ROC曲线可对图像质量的主观评价进行量化分析。
关键词:  ROC曲线  周期性旋转重叠平行采集和增强重建  弥散加权成像  磁敏感伪影
DOI:10.3724/SP.J.1008.2012.001383
投稿时间:2012-10-20修订日期:2012-11-10
基金项目:
Performance of PROPELLER technique in reducing susceptibility artifacts in diffusion-weighted MR images
LIU Xiao-yan1,LIU Ting-ting1,WANG Jun2*
(1. Department of Radiology, Affiliated Hospital of Nantong University, Nantong 226001, Jiangsu, China
2. Department of Radiology, Nanjing General Hospital, PLA Nanjing Military Area Command, Nanjing 210002, Jiangsu, China
*Corresponding author.)
Abstract:
ObjectiveTo evaluate the performance of PROPELLER diffusion-weighted magnetic resonance image (DWI) in reducing image distortion caused by susceptibility artifacts. MethodsForty volunteers with metal denture were enrolled in the present study; their PROPELLER and echo-planar DWI were taken for the skull base. The quality of each image was evaluated by using a four-point scoring system. Qualitative comparison of image distortion between PROPELLER DWI and echo-planar DWI was performed by ROC analysis. ResultsThe area under the ROC curve was 0.979 (P<0.01), indicating a significant difference in the image distortion between PROPELLER DWI and echo-planar DWI. PROPELLER technique could greatly reduce the image distortion associated with magnetic susceptibility artifacts. ConclusionPROPELLER technique can greatly reduce metal denture-induced susceptibility-related image distortion, and ROC curve can be used for quantitative analysis of subjective assessment of image quality.
Key words:  ROC curve  periodically rotated overlapping parallel lines with enhanced reconstruction  diffusion-weighted image  magnetic susceptibility artifacts